Jul 9, 2017 19:43
6 yrs ago
English term
by discharge to earth
English to Russian
Medical
Medical: Dentistry
инструкция к хирургическому скалеру
Technical data:
Protection by discharge to earth: 10 ms
Protection by discharge to earth: 10 ms
Proposed translations
(Russian)
4 +1 | см. | Natalie |
4 | * | hawkwind |
3 | посредством/путём заземления | Oleg Lozinskiy |
3 | путем разряда на землю | mk_lab |
Proposed translations
+1
28 mins
Selected
см.
Защита путем сброса импульса тока на "землю": время сброса 10 мс
4 KudoZ points awarded for this answer.
Comment: "Thx!"
7 mins
посредством/путём заземления
*
--------------------------------------------------
Note added at 15 мин (2017-07-09 19:59:46 GMT)
--------------------------------------------------
IMHO, 'discharge to earth' = 'earthing' = 'grounding', etc.
--------------------------------------------------
Note added at 20 мин (2017-07-09 20:04:46 GMT)
--------------------------------------------------
P.S. Подробнее можно прочитать здесь: https://ru.wikipedia.org/wiki/Заземление
--------------------------------------------------
Note added at 15 мин (2017-07-09 19:59:46 GMT)
--------------------------------------------------
IMHO, 'discharge to earth' = 'earthing' = 'grounding', etc.
--------------------------------------------------
Note added at 20 мин (2017-07-09 20:04:46 GMT)
--------------------------------------------------
P.S. Подробнее можно прочитать здесь: https://ru.wikipedia.org/wiki/Заземление
59 mins
путем разряда на землю
или "сброса (электростатического) заряде на землю"
По-видимому, речь идет о периодическом сбросе электростатического заряда, формирующегося в процессе трения на скалере, чтобы разряд не причинил боль пациенту. Для этого инструмент периодически кратковременно (на 10 мс) заземляется - разряжается на землю. Это делается автоматически встроенной электроникой
По-видимому, речь идет о периодическом сбросе электростатического заряда, формирующегося в процессе трения на скалере, чтобы разряд не причинил боль пациенту. Для этого инструмент периодически кратковременно (на 10 мс) заземляется - разряжается на землю. Это делается автоматически встроенной электроникой
18 hrs
*
время срабатывания автоматической защиты при пробое на землю
http://pdf.medicalexpo.com/pdf/mectron-spa/piezo-smart/73174...
Automatic Protection Control (APC)
15.8 Protection systems and tripping time of the APC:
No handpiece connected: 10ms
Cord interrupted: 10ms
Tips broken or not correctly tightened: <500ms
Protection by discharge to earth: 10ms
http://www.usdentaldepot.com/usdd/2/usdd_scaler.pdf
* Automatic Protection Control: interrupts power and water supply in case of using incorrect handpiece and insert tip..
--------------------------------------------------
Note added at 1 day4 hrs (2017-07-11 00:27:41 GMT)
--------------------------------------------------
http://www.anas.ru/linked/files/attach/Instruktsiya_Compact_...
--------------------------------------------------
Note added at 1 day4 hrs (2017-07-11 00:29:54 GMT)
--------------------------------------------------
Автоматическая система защиты
ACЗ гарантирует максимальную безопасность. Система автоматически распознает любые отклонения от нормальной работы piezo smart и блокирует генерацию ультразвука и подачу жидкости менее чем через 0,1 с. Например, если причина остановки неправильное подсоединение насадки, то она будет выведена на дисплей в виде кода ошибки.
--------------------------------------------------
Note added at 1 day5 hrs (2017-07-11 01:23:07 GMT)
--------------------------------------------------
NOTE: APC activation.
The device is equipped with an APC (automatic protection circuit), which is activated when any
upset occurs during operations. APC activation is indicated by the yellow LED on the front of
the device (Fig.5 - Ref.G). According to the reason of the upset, two kind of alarms are possible:
flashing or fixed light, as follows:
flashing check LED:
- Insert not properly fitted in/screwed onto handpiece;
- Insert worn out, broken or deformed;
- Handpiece electric contacts and/or cord connectors are wet.
Т.е. одна из причин - мокрые контакты и высокое напряжение генератора, подаваемое на пьезокристалл будет "шить" на землю кабеля
--------------------------------------------------
Note added at 1 day5 hrs (2017-07-11 01:26:16 GMT)
--------------------------------------------------
WARNING: The electric contacts of the cord connector must be dry.
At the end of the sterilisation cycle, before connecting the cord to the device, make sure that the
electrical contacts of the connector are completely dry, if necessary dry the contacts by blowing
compressed air with the syringe.
--------------------------------------------------
Note added at 1 day5 hrs (2017-07-11 01:30:09 GMT)
--------------------------------------------------
И здесь несколько по-другому уточнено:
APC activation: - No handpiece
- Discontinuity of cord wire
- Insert not properly tightened or broken
- Intervention of the ground discharge protection
system
--------------------------------------------------
Note added at 1 day5 hrs (2017-07-11 01:30:23 GMT)
--------------------------------------------------
http://dental.mectron.com/fileadmin/user_upload/dental/engli...
http://pdf.medicalexpo.com/pdf/mectron-spa/piezo-smart/73174...
Automatic Protection Control (APC)
15.8 Protection systems and tripping time of the APC:
No handpiece connected: 10ms
Cord interrupted: 10ms
Tips broken or not correctly tightened: <500ms
Protection by discharge to earth: 10ms
http://www.usdentaldepot.com/usdd/2/usdd_scaler.pdf
* Automatic Protection Control: interrupts power and water supply in case of using incorrect handpiece and insert tip..
--------------------------------------------------
Note added at 1 day4 hrs (2017-07-11 00:27:41 GMT)
--------------------------------------------------
http://www.anas.ru/linked/files/attach/Instruktsiya_Compact_...
--------------------------------------------------
Note added at 1 day4 hrs (2017-07-11 00:29:54 GMT)
--------------------------------------------------
Автоматическая система защиты
ACЗ гарантирует максимальную безопасность. Система автоматически распознает любые отклонения от нормальной работы piezo smart и блокирует генерацию ультразвука и подачу жидкости менее чем через 0,1 с. Например, если причина остановки неправильное подсоединение насадки, то она будет выведена на дисплей в виде кода ошибки.
--------------------------------------------------
Note added at 1 day5 hrs (2017-07-11 01:23:07 GMT)
--------------------------------------------------
NOTE: APC activation.
The device is equipped with an APC (automatic protection circuit), which is activated when any
upset occurs during operations. APC activation is indicated by the yellow LED on the front of
the device (Fig.5 - Ref.G). According to the reason of the upset, two kind of alarms are possible:
flashing or fixed light, as follows:
flashing check LED:
- Insert not properly fitted in/screwed onto handpiece;
- Insert worn out, broken or deformed;
- Handpiece electric contacts and/or cord connectors are wet.
Т.е. одна из причин - мокрые контакты и высокое напряжение генератора, подаваемое на пьезокристалл будет "шить" на землю кабеля
--------------------------------------------------
Note added at 1 day5 hrs (2017-07-11 01:26:16 GMT)
--------------------------------------------------
WARNING: The electric contacts of the cord connector must be dry.
At the end of the sterilisation cycle, before connecting the cord to the device, make sure that the
electrical contacts of the connector are completely dry, if necessary dry the contacts by blowing
compressed air with the syringe.
--------------------------------------------------
Note added at 1 day5 hrs (2017-07-11 01:30:09 GMT)
--------------------------------------------------
И здесь несколько по-другому уточнено:
APC activation: - No handpiece
- Discontinuity of cord wire
- Insert not properly tightened or broken
- Intervention of the ground discharge protection
system
--------------------------------------------------
Note added at 1 day5 hrs (2017-07-11 01:30:23 GMT)
--------------------------------------------------
http://dental.mectron.com/fileadmin/user_upload/dental/engli...
Something went wrong...